Hygiene Monitoring
Designed for efficacy validation of Chemdye® PRO1 MICRO Hygiene Monitoring System.


Product type
PRO1 VT Chemdye® Validation Kit for efficacy validation of the PRO1 MICRO Hygiene Monitoring System
Main application
Validates the performance and sensitivity of the PRO1 MICRO Hygiene Monitoring System when used with Bionova® MiniPro, IC10/20FR, or IC10/20FRLCD Auto‑Reader Incubators
Key benefit
Ensures accurate detection of residual proteins and reducing agents on surfaces and hard‑to‑reach areas after cleaning, by providing standardized positive and negative controls
Technology / differentiator
- Includes Negative Control vial (protein‑free solution, transparent cap) and Positive Control vial (defined protein standard, light blue cap) - Vial dimensions: 12 mm diameter × 47 mm length - Quantitative analysis performed with Bionova® Auto‑Reader incubators - Incubation time: 7 minutes (IC10/20FR / IC10/20FRLCD) or 4 minutes (MiniPro) - Incubation range: 60 ± 2 °C - Designed under ISO 13485:2016, ISO 15883‑1:2006, HTM 01‑01:2016, HTM 01‑05:2013 standards
Compatibility / usage environment
Exclusively for validation of the PRO1 MICRO Hygiene Monitoring System; suitable for healthcare, food, pharmaceutical, and medical industries