Skip to content

Chemical indicator for Hydrogen Peroxide (FDA)

SKU: CD40 Category: Tags , ,
Additional information
Brand

Chemdye

Process

Hydrogen Peroxide

Packaging

250

Initial Color

Purple

Final Color

Green

Indicator Type

Type 4 Chemical Indicator

Regulations

ISO 11140-1

Description

CD40 / Multi-variable Chemical Indicator, strips printed with reactive ink for Plasma or Vaporized Hydrogen Peroxide sterilization processes.

Control of plasma or vaporized H2O2 sterilization processes. Conditions: 6 minutes, 50 °C, 2,3 mg/l H2O2.

Please review the instructions for use for applicable cycles in your country

Characteristics

-Type 4 Chemical Indicator (according to ISO 11140-1).

-Strips printed with indicator ink.

-Size: 105 x 18 mm (approximate).

-Initial color: Purple.

-Final color: Green.

-100 % Toxic Heavy Metals free.

Instructions for Use

1.Place a Chemdye® CD40 Multivariable Process Indicator within each package, bag or tray that will be sterilized by Plasma or Vaporized Hydrogen Peroxide. In rigid containers, place the indicator in every corner, or at least two diagonally opposite corners, to ensure that the sterilization agent has reached the most inaccessible areas.

2.Perform the sterilization cycle.

3.After the sterilization process has finished, remove the indicators from the sterilizer and analyze the results. For all the chemical indicators exposed to Plasma or Vaporized Hydrogen Peroxide, the ink must turn to the green or lighter reference color when the sterilization parameters are reached.

Description

CD40 / Multi-variable Chemical Indicator, strips printed with reactive ink for Plasma or Vaporized Hydrogen Peroxide sterilization processes.

Control of plasma or vaporized H2O2 sterilization processes. Conditions: 6 minutes, 50 °C, 2,3 mg/l H2O2.

Please review the instructions for use for applicable cycles in your country

Characteristics

-Type 4 Chemical Indicator (according to ISO 11140-1).

-Strips printed with indicator ink.

-Size: 105 x 18 mm (approximate).

-Initial color: Purple.

-Final color: Green.

-100 % Toxic Heavy Metals free.

Instructions for Use

1.Place a Chemdye® CD40 Multivariable Process Indicator within each package, bag or tray that will be sterilized by Plasma or Vaporized Hydrogen Peroxide. In rigid containers, place the indicator in every corner, or at least two diagonally opposite corners, to ensure that the sterilization agent has reached the most inaccessible areas.

2.Perform the sterilization cycle.

3.After the sterilization process has finished, remove the indicators from the sterilizer and analyze the results. For all the chemical indicators exposed to Plasma or Vaporized Hydrogen Peroxide, the ink must turn to the green or lighter reference color when the sterilization parameters are reached.