Chemical indicator Liquid for Peracetic Acid

Additional information
Brand

Integron

Process

Endoscope Reprocessing, PAA

Packaging

100

Initial Color

Purple

Final Color

Fuchsia

Regulations

IRAM 37102-1, ISO 11140-1

Description

Package of non-absorbent material (25 mm x 70 mm). Polypropylene strip (10 mm x 45 mm) printed with indicator ink for monitoring of High Level Disinfection processes using Liquid Peracetic Acid. Turns from Violet to Fuchsia.

Integron® IT400 Indicators are designed to react in High Level Disinfection processes that use Peracetic Acid (PAA) as disinfecting agent, assuring an adequate control of the disinfection process efficacy (temperature, time and PAA concentration).

The reactive ink was developed for processes that work at temperatures between 50-60 °C and turns to fuchsia when integration conditions reached the established values stated in the quality certificate.

Instructions of use

1. With the reusable silicon strip provided along with the product, place the envelope with the chemical test strip inside the equipment, together with the load to be processed with PAA.

2. Process the load according to the established procedures.

3. After completing the process, analyze the results. The indicator must change to the reference color for considering that the desired disinfection conditions were met.

Description

Package of non-absorbent material (25 mm x 70 mm). Polypropylene strip (10 mm x 45 mm) printed with indicator ink for monitoring of High Level Disinfection processes using Liquid Peracetic Acid. Turns from Violet to Fuchsia.

Integron® IT400 Indicators are designed to react in High Level Disinfection processes that use Peracetic Acid (PAA) as disinfecting agent, assuring an adequate control of the disinfection process efficacy (temperature, time and PAA concentration).

The reactive ink was developed for processes that work at temperatures between 50-60 °C and turns to fuchsia when integration conditions reached the established values stated in the quality certificate.

Instructions of use

1. With the reusable silicon strip provided along with the product, place the envelope with the chemical test strip inside the equipment, together with the load to be processed with PAA.

2. Process the load according to the established procedures.

3. After completing the process, analyze the results. The indicator must change to the reference color for considering that the desired disinfection conditions were met.